Product
 
   
   
   
   
   
 
   
   
   
   
   
   
     
     
Prevention Therapy
English - Español - Français - Deutsch
Novacare.de
Novacare
»
 
 
Technical Data
Article #: 991500N

Control unit
Measurements: 390 x 220 x 121 mm Weight: 5.5 kg

2 membrane compressors, digital control with pressure sensor, alternating pressure, staticmode, care mode, 3 time cycles (5/10/15 min.), automatic lock pad, low pressure alarm, power loss alarm

Mattress
Measurements, inflated incl. base mattress: appr. 200 x 90 x 24 cm weight: appr. 11.8 kg

20 cells, 3 static non-bending connecting hoses, CPR emergency ventilation (incl. transport cap)

Materials
Cells: nylon with polyurethane coating Cover: Bi-elastic cover with Polyurethane/Polyester coating, water-repellent, breathable Base: nylon with integrated pocket for 3.5 cm base mattress

Accessories Including transport bag for matress and control unit

Max. patient weight : 180 kg
APM 440 plus
APM 440 plus APM 440 plus
The APM 440 alternating pressure system is suited for the treatment of
decubitus up to stage IV.

The control unit is equipped with a digital sensor control, which regulates the inner pressure of the mattress, creating a perfect surface pressure. Again, the use of two compressors guarantees safety in case of a compressor malfunction.

Adjustment is accomplished with the aid of a touch pad surface, equipped with visually easy to comprehend pictograms. The result is a quick and easy operation.

The alternating mode offers three separate time cycles. In addition a permanent (pressure reduced) static mode can be selected for the therapy of pain patients, as well as a care mode, which simplifies care and mobilization. Once in care mode all mattress cells are inflated for 30 minutes. This significantly simplifies the re-positioning and the care of patients.erleichtert. The automatic key pad lock is activated after three minutes preventing an unintentional “re-adjustment” of the control unit.

The mattress cells are 24 cm high, this allows the patient to sink deeper into the matress surface. As a result patients weighing up to 180 kg can be positioned. Each cell is equipped with a CPC quick coupling, permitting a cell exchange within seconds or cell removal to freely position the patient or for cleaning.

Our transport securing device maintains the system inflated, thus guaranteeing a safe transport of patients. As an additional safety measure our mattress substitution system is equipped with a 3.5 cm high base mattress, for insertion into an integrated pocket.


Not Available For Purchase

Model
Attributes
991200+
991300N
991400N
991600N
991600N